Subsampling Error in Powders: How to Qualify Your Splitting Procedure
Qualify powder splitting with independent preparation paths, variance components, reference comparisons, and acceptance criteria suited to the intended measurement.
Qualify powder splitting with independent preparation paths, variance components, reference comparisons, and acceptance criteria suited to the intended measurement.
A sample-cup film change can alter detected XRF fluorescence. Check film attenuation, calibration compatibility, blanks, and sample presentation before treating the reported shift as a powder composition change.
Changed XRD peak ratios after repacking can reflect crystallite alignment or poor particle statistics. Compare independent mounts before interpreting phase conversion.
Moisture content analysis in powders is a critical aspect [...]
Bulk density measurement, as a key aspect of material [...]